Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
57
Moored Mode Setup Commands
Moored mode setup commands following MM apply to Moored Mode
only, and have no effect on operation if the 19plus V2 is in Profiling mode.
MM Set 19plus V2 to Moored mode.
SampleInterval=x x= interval (seconds) between samples
(10 - 14,400 seconds).
NCycles=x x= number of measurements to take and
average every SampleInterval seconds.
Range 1 – 100; default 1. 19plus V2 takes
and averages NCycles samples (each
0.25 seconds apart) each SampleInterval
seconds; averaged data is stored in FLASH
memory and (if MooredTxRealTime=Y)
transmitted real-time.
MooredPumpMode=x x=0: No pump.
x=1: Run pump for 0.5 seconds before
each sample (typical for pumping through
conductivity cell only, with no auxiliary
sensors connected to plumbing).
x=2: Run pump during each sample
(typical for pumping through conductivity
cell and in-line auxiliary sensor).
DelayBeforeSampling=x x= time (seconds) to wait after switching
on external voltage and RS-232 sensors
before sampling (0-600 seconds). Default
0 seconds. Typical values if using:
- SBE 43 oxygen sensor – time is
dependent on membrane thickness and
on water temperature (see Pump
Operation – Moored Mode) . Use with
MooredPumpMode=2.
- Beckman- or YSI-type oxygen sensor –
120 to 180 seconds required to provide
time for sensor to polarize.
Use with MooredPumpMode=2.
- Sea Tech fluorometer - 15 seconds
required to provide time for sensor
to stabilize.
MooredTxRealTime=x x=Y: Output real-time data.
x=N: Do not.
• Pump operation is affected by both
DelayBeforeSampling= and
MooredPumpMode=. See Pump
Operation – Moored Mode.
• NCycles=, DelayBeforeSampling=,
MooredPumpMode=, and
MooredTxRealTime= affect the
time required to sample. If the time
required to sample is too high, the
19plus V2 is unable to take the
required number of measurements
and do the calculations within
SampleInterval=. See Sample
Timing in Section 2: Description of
SBE 19plus V2.
• Power turn-on defined by
DelayBeforeSampling= does not
apply to the RS-232 Aanderaa
Optode, because turning on power
to the Optode causes it to sample.
The 19plus V2 turns on power to the
Optode when the 19plus V2 is ready
to acquire the Optode data.
The 19plus V2 requires verification
when MM is sent. 19plus V2 responds:
this command will change the
scan length and/or initialize
logging. Repeat the command to
verify.
Type the command again
and press the Enter key to proceed.
The 19plus V2 responds:
Scan length has changed,
Example: If SampleInterval=20 and NCycles=4, every 20 sec 19plus V2 takes 4 measurements
(0.25 sec apart), averages data from 4 measurements, and stores averaged data in FLASH memory.
• For 19plus V2 with optional Quartz
pressure sensor, NCycles= must be
an even number (2, 4, etc.);
increasing NCycles= reduces
measurement noise. See
Specifications in Section 2:
Description of SBE 19plus V2 to
determine resolution.
• 19plus V2 integrates for the Quartz
pressure sensor after NCycles=
measurements have been taken.