Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
31
Profiling Mode
The SBE 19plus V2 samples data at 4 Hz (one sample every 0.25 seconds),
averages the data at pre-programmed intervals, stores the averaged data in its
FLASH memory, and transmits the averaged data real-time. The 19plus V2
provides several methods for starting and stopping logging, depending on the
settings for IgnoreSwitch= and AutoRun=:
IgnoreSwitch= AutoRun=
N N
Slide magnetic
switch on.
Slide magnetic switch off, or
send
Stop
Y N
StartDateTime=
StartLater
Send Stop.
Y or N Y
Turn on external
power.
• Turn off external power, or
• (if you want to send
commands to check or
Stop.
Example: 19plus V2 in Profiling mode
Wake up 19plus V2. Set date and time to October 1, 2011 at 9:05 am. Initialize
logging to overwrite previous data in memory. Set up with 1 voltage sensor,
average every 4 samples, and output data in raw hex format. Set up with a 60-
second pump turn-on delay after pump enters water, to ensure pump is primed
before turning on. Set up to initiate logging with the magnetic switch. After all
parameters are entered, verify setup with status command. Send power-off
command.
(Click Connect in Seaterm232’s Communications menu to connect and wake up.)
DATETIME=10012011090500
INITLOGGING
VOLT0=Y
NAVG=4
OUTPUTFORMAT=0
PUMPDELAY=60
IGNORESWITCH=N
GETCD (to verify setup)
QS
Start logging by putting magnetic switch in On position. Put 19plus V2 in water,
and allow to soak for at least time required for pump turn-on (PumpDelay=60)
before beginning downcast. If desired, use Seasave to view real-time data. When
cast is complete, stop logging by putting magnetic switch in Off position.
Upload data in memory, in format SBE Data Processing and Seasave can use.
Send power-off command.
(Click Connect in Seaterm232’s Communications menu to connect and wake up.)
(Click Upload– Seaterm232 leads you through screens to define data to be
uploaded and where to store it.)
QS
Sea-Bird ships the 19plus V2
with AutoRun=N (it will not
automatically start sampling
when external power is
applied). If you send
AutoRun=Y:
• Send QS to put 19plus V2 in
quiescent (sleep) state, and
then turn power off and then
on again to start sampling.
or
•
Send StartNow to
start sampling.
Use Stop to:
• Stop sampling
• Stop waiting to start sampling (after
Startlater has been sent)
Once Stop is sent, the 19plus V2
accepts all commands again.