Manual revision 006 Appendix VI: Manual Revision History SBE 19plus V2
122
006
04/11
Firmware 2.3:
Add compatibility with WET Labs RS-232 sensors (WetLabs=, TWetLabs).
Add commands for RS-232 sensor type and serial number (SetSerialType=, SetSerialSN=).
StartLater limitation is 1 month in future.
Correct explanation of SampleNumber=; does not reset HeaderNumber=.
Clarify use of DelayBeforeSampling=; applies to all auxiliary sensors except Optode.
Clarify how to stop logging if in profiling mode and IgnoreSwitch=N; must use switch or send
<CTRL Z>; Stop command only stops logging briefly, and then logging resumes.
Correct Digiquartz pressure sensor stability specification to 0.02% of full scale range/year.
Clarify that in Moored mode, Optode data is acquired after all CTD and other external sensor
data.
Replacement parts list: add ice cage, update Seaspares kit contents
Update SeatermV2 screen.
Remove references to Druck pressure sensors.
Add information that Ni-Cad battery pack and charger no longer available.
Correct typos.