SBE 19plus V2
SEACAT Profiler
Conductivity, Temperature, and Pressure Recorder
with RS-232 Interface
Note: NEW ADDRESS
As of January 2010
User’s Manual
Sea-Bird Electronics, Inc.
13431 NE 20
th
Street
Bellevue, Washington 98005 USA Manual Version #006, 04/21/11
Telephone: +1 425-643-9866 Firmware Version 2.3 and later
Fax: +1 425-643-9954 SeatermV2 Version 1.1e and later
[email protected] Seasave V7 Version 7.21d and later
www.seabird.com SBE Data Processing Version 7.21d and later
Zobrazit stránku 0
1 2 3 4 5 6 ... 123 124

Shrnutí obsahu

Strany 1 - SEACAT Profiler

SBE 19plus V2 SEACAT Profiler Conductivity, Temperature, and Pressure Recorder with RS-232 Interface Note: NEW ADDRESS As of Ja

Strany 2 - Limited Liability Statement

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 10 • SBE 55 ECO Water Sampler - The SBE 55 has built-in capabilities simi

Strany 3 - Table of Contents

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 100 Replacing Anti-Foulant Devices (SBE 16plus, SBE 19plus) The S

Strany 4

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 101 Sensor Calibration Sea-Bird sensors are calibrated by subjecti

Strany 5 - Section 1: Introduction

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 102 Pressure Sensor The SBE 19plus V2 includes a standard strain-g

Strany 6 - Unpacking SBE 19plus V2

Manual revision 006 Section 6: Troubleshooting SBE 19plus V2 103 Section 6: Troubleshooting This section reviews common problems in operating the S

Strany 7 - System Description

Manual revision 006 Section 6: Troubleshooting SBE 19plus V2 104 Problem 3: Nonsense or Unreasonable Data The symptom of this problem is a data fil

Strany 8

Manual revision 006 Glossary SBE 19plus V2 105 Glossary Batteries – Nine alkaline D-cells (Duracell MN1300, LR20) standard. Available with optional

Strany 9

Manual revision 006 Glossary SBE 19plus V2 106 Seaterm232 – Win 2000/XP terminal program used with Sea-Bird instruments that communicate via an RS-2

Strany 10

Manual revision 006 Appendix I: Functional Description and Circuitry SBE 19plus V2 107 Appendix I: Functional Description and Circuitry Sensors T

Strany 11 - Specifications

Manual revision 006 Appendix I: Functional Description and Circuitry SBE 19plus V2 108 Battery Wiring SBE 19plus V2 standard main battery is a seri

Strany 12 - Continued from previous page

Manual revision 006 Appendix II: Electronics Disassembly/Reassembly SBE 19plus V2 109 Appendix II: Electronics Disassembly/Reassembly Disassem

Strany 13

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 11 Specifications Temperature (°C) Conductivity (S/m) Pressure Measurem

Strany 14 - Sample Timing (Moored Mode)

Manual revision 006 Appendix III: Command Summary SBE 19plus V2 110 Appendix III: Command Summary CATEGORY COMMAND DESCRIPTION Status GetCD Get an

Strany 15 - Battery Endurance

Manual revision 006 Appendix III: Command Summary SBE 19plus V2 111 CATEGORY COMMAND DESCRIPTION Output Format Setup OutputFormat=x x=0: output r

Strany 16

Manual revision 006 Appendix III: Command Summary SBE 19plus V2 112 CATEGORY COMMAND DESCRIPTION Data Upload Stop logging before uploading. GetSam

Strany 17 - External Power

Manual revision 006 Appendix III: Command Summary SBE 19plus V2 113 CATEGORY COMMAND DESCRIPTION Calibration Coefficients (F=floating point number

Strany 18

Manual revision 006 Appendix IV: AF24173 Anti-Foulant Device SBE 19plus V2 114 Appendix IV: AF24173 Anti-Foulant Device AF24173 Anti-Foulant Devic

Strany 19 - Magnetic Reed Switch

Manual revision 006 Appendix IV: AF24173 Anti-Foulant Device SBE 19plus V2 115 AF24173 Anti-Foulant Device FOR USE ONLY IN SEA-BIRD ELECTRONICS’ CO

Strany 20 - Vertical Mount

Manual revision 006 Appendix IV: AF24173 Anti-Foulant Device SBE 19plus V2 116 PRECAUTIONARY STATEMENTS HAZARD TO HUMANS AND DOMESTIC ANIMALS DANG

Strany 21

Manual revision 006 Appendix IV: AF24173 Anti-Foulant Device SBE 19plus V2 117 STORAGE AND DISPOSAL PESTICIDE STORAGE: Store in original containe

Strany 22

Manual revision 006 Appendix V: Replacement Parts SBE 19plus V2 118 Appendix V: Replacement Parts Part Number Part Application Description Quantit

Strany 23 - Horizontal Mount

Manual revision 006 Appendix V: Replacement Parts SBE 19plus V2 119 continued from previous page Part Number Part Application Description Quantit

Strany 24 - Section 3:

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 12 Continued from previous page Internal Batteries (nominal capacity) Sta

Strany 25 - See Seaterm232’s Help files

Manual revision 006 Appendix V: Replacement Parts SBE 19plus V2 120 continued from previous page Part Number Part Application Description Quantit

Strany 26

Manual revision 006 Appendix VI: Manual Revision History SBE 19plus V2 121 Appendix VI: Manual Revision History Manual Version Date Description 00

Strany 27

Manual revision 006 Appendix VI: Manual Revision History SBE 19plus V2 122 006 04/11 • Firmware 2.3: Add compatibility with WET Labs RS-232 senso

Strany 28

Manual revision 006 Index SBE 19plus V2 123 Index . .con file · 50, 51, 64, 78, 80, 89, 101 .xmlcon file · 50, 51, 64, 78, 80, 89, 101 A Aanderaa Op

Strany 29

Manual revision 006 Index SBE 19plus V2 124 M Magnetic reed switch · 19 Maintenance · 92 Manual revision history · 121 Memory · 11, 19 Modes · See S

Strany 30

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 13 Dimensions and End Cap Connectors Dimensions in millimeters (inches)

Strany 31 - Profiling Mode

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 14 Sample Timing (Moored Mode) Sample timing in Moored Mode is dependent

Strany 32 - Moored Mode

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 15 Batteries For the main battery, the SBE 19plus V2 uses nine D-cell al

Strany 33 - Pump Operation - General

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 16 Moored Mode Examples A 19plus V2 with strain-gauge pressure sensor a

Strany 34 - Pump Operation - Moored Mode

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 17 External Power The SBE 19plus V2 can be powered from an external sourc

Strany 35

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 18 Supplying Enough Power to SBE 19plus V2 Another consideration in de

Strany 36 - Real-Time Setup

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 19 Data Storage The SBE 19plus V2 has a 64 Mbyte FLASH memory. Shown bel

Strany 37 - Timeout Description

2 Limited Liability Statement Extreme care should be exercised when using or servicing this equipment. It should be used or serviced only by personn

Strany 38 - Command Descriptions

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 20 Configuration Options and Plumbing The standard SBE 19plus V2 include

Strany 39

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 21 Shown below is the vertical mount plumbing arrangement of a 19plus V2

Strany 40 - Status Commands (continued)

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 22 Shown below is the vertical mount plumbing arrangement of a 19plus V2 c

Strany 41

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 23 Horizontal Mount Shown below is the horizontal mount plumbing arrangem

Strany 42

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 24 Section 3: Power and Communications Test This section describes soft

Strany 43

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 25 Test 1. Double click on SeatermV2.exe. The main screen looks like th

Strany 44

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 26 Following is a description of the menus: Menu Description Equivalent

Strany 45

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 27 3. If this is the first time Seaterm232 is being used, the Serial Por

Strany 46

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 28 Taking a look at the Send Commands window: You can use the Send Com

Strany 47

Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2 29 5. Display 19plus V2 status information by typing DS and pressing th

Strany 48

Manual revision 006 Table of Contents SBE 19plus V2 3 Table of Contents Section 1: Introduction ...

Strany 49

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 30 Section 4: Deploying and Operating SBE 19plus V2 This sectio

Strany 50

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 31 Profiling Mode The SBE 19plus V2 samples data at 4 Hz (one sa

Strany 51

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 32 Moored Mode At pre-programmed intervals, the SBE 19plus V2 wa

Strany 52

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 33 Pump Operation - General Do not run the pump dry. The pump is

Strany 53

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 34 Pump Operation - Moored Mode Pump operation is governed by tw

Strany 54

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 35 Moored Mode Pump Setting Recommendations Sea-Bird provides th

Strany 55

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 36 Real-Time Setup Baud Rate and Cable Length Without a Sea-Bi

Strany 56

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 37 With a Sea-Bird Deck Unit Set the 19plus V2 baud rate to 4800

Strany 57

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 38 Command Descriptions This section describes commands and prov

Strany 58 - Moored Mode (MM)

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 39 Status Commands GetCD Get and display configuration data, wh

Strany 59

Manual revision 006 Table of Contents SBE 19plus V2 4 Section 5: Routine Maintenance and Calibration ... 92 Corrosion Precau

Strany 60

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 40 Status Commands (continued) Example: 19plus V2 in Profiling

Strany 61

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 41 Status Commands (continued) GetSD Get and display status dat

Strany 62

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 42 Status Commands (continued) GetCC Get and display calibratio

Strany 63

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 43 Status Commands (continued) GetEC Get and display event coun

Strany 64

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 44 Status Commands (continued) GetHD Get and display hardware d

Strany 65

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 45 Status Commands (continued) DS Get and display operating stat

Strany 66 - Data Output Formats

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 46 Status Commands (continued) Moored Mode (MM) • Firmware v

Strany 67

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 47 Status Commands (continued) DCal Get and display calibration

Strany 68

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 48 General Setup Commands DateTime=mmddyyyyhhmmss Set real-tim

Strany 69

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 49 General Setup Commands (continued) InitLogging Initialize lo

Strany 70

Manual revision 006 Section 1: Introduction SBE 19plus V2 5 Section 1: Introduction This section includes a Quick Start procedure, and photos of a

Strany 71

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 50 Pressure and Voltage Sensor Setup Commands PType=x Pressur

Strany 72

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 51 RS-232 Sensor Setup Commands The SBE 19plus V2 can interface

Strany 73

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 52 RS-232 Sensor Setup Commands (continued) Setup for WET Labs S

Strany 74

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 53 RS-232 Sensor Setup Commands (continued) Setup for GTD: SBE

Strany 75

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 54 RS-232 Sensor Setup Commands (continued) Setup for Aanderaa

Strany 76

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 55 Output Format Setup Commands See Data Output Formats after t

Strany 77

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 56 Profiling Mode Setup Commands Profiling mode setup commands

Strany 78 - Setup for Deployment

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 57 Moored Mode Setup Commands Moored mode setup commands follow

Strany 79 - Deployment

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 58 Logging Commands Logging commands direct the SBE 19plus V2 t

Strany 80

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 59 Logging Commands (continued) StartNow Start logging now.

Strany 81

Manual revision 006 Section 1: Introduction SBE 19plus V2 6 Unpacking SBE 19plus V2 Shown below is a typical SBE 19plus V2 shipment.

Strany 82 - Acquiring Real-Time Data

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 60 Data Upload Commands Stop logging before uploading data. If m

Strany 83

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 61 Polled Sampling Commands These commands request a single samp

Strany 84 - Recovery

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 62 Testing Commands The 19plus V2 takes and outputs 100 samples

Strany 85 - Uploading Data

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 63 Testing Commands continued TWetLabs Measure WET Labs RS-232

Strany 86

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 64 Calibration Coefficients Commands Calibration coefficients ar

Strany 87 - Command/Data Echo Area

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 65 Hardware Configuration Commands The following commands are us

Strany 88

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 66 Data Output Formats The SBE 19plus V2 stores data in a compac

Strany 89

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 67 OutputFormat=0 (raw frequencies and voltages in Hex) Data is

Strany 90

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 68 OutputFormat=1 (engineering units in Hex) Data is output in t

Strany 91 - Editing Raw Data File

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 69 OutputFormat=2 (raw frequencies and voltages in decimal) Data

Strany 92 - Corrosion Precautions

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 7 Section 2: Description of SBE 19plus V2 This section describes the func

Strany 93 - Plumbing Maintenance

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 70 OutputFormat=3 (engineering units in decimal) Data is output

Strany 94 - Replacing Alkaline Batteries

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 71 OutputFormat=4 (pressure and scan number in Hex) Data is outp

Strany 95

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 72 Uploaded Data (from GetSamples:b,e or DDb,e): <?xml?> &l

Strany 96

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 73 Polled Data from SL, SLT, TS, TSS, or TSSOn: If OutputUCSD=Y a

Strany 97 - Do not recharge the Ni-Cad

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 74 Optimizing Data Quality for Profiling Applications Our deploy

Strany 98 - Barbed caps for cleaning

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 75 When very heavy seas cause severe ship motion, the 19plus V2 d

Strany 99 - Pump Maintenance

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 76 Installing Anti-Foul Fittings for Moored Applications The SBE

Strany 100 - 8-inch socket wrench;

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 77 4. Intake – A. Remove the two hex head screws attaching the

Strany 101 - Sensor Calibration

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 78 Setup for Deployment 1. Install new batteries or ensure the

Strany 102 - Pressure Sensor

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 79 Deployment 1. Install a cable or dummy plug for each connec

Strany 103 - Section 6: Troubleshooting

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 8 SBE 19plus V2 options / accessories include: • Titanium housing for use

Strany 104 - Problem 4: Program Corrupted

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 80 Acquiring Real-Time Data with Seasave Verifying Contents of C

Strany 105 - Glossary

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 81 3. The configuration information appears on the Instrument C

Strany 106

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 82 Acquiring Real-Time Data Instructions below are for an SBE 19

Strany 107 - Sensors

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 83 5. In the Start Real-Time Data Acquisition dialog box, click

Strany 108 - Battery Wiring

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 84 Recovery Physical Handling 1. Rinse the instrument and cond

Strany 109 - Appendix II:

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 85 Uploading Data 1. Double click on SeatermV2.exe. The main sc

Strany 110 - WetLabs=x

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 86 6. Click Upload to upload stored data. Seaterm232 responds as

Strany 111

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 87 7. Click the Header Form tab to customize the header: The ent

Strany 112

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 88 9. Ensure all data has been uploaded by reviewing and proces

Strany 113

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 89 Processing Data Using SBE Data Processing 1. Convert the .he

Strany 114 - AF24173 ANTI-FOULANT DEVICE

Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2 9 • SBE 36 CTD Deck Unit and Power Data Interface Module (PDIM) - These i

Strany 115 - AF24173 Anti-Foulant Device

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 90 3. In the Open dialog box, select the .xmlcon or .con file an

Strany 116

Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2 91 Editing Raw Data File Sometimes users want to edit the raw .h

Strany 117

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 92 Section 5: Routine Maintenance and Calibration This section re

Strany 118

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 93 Plumbing Maintenance For an SBE 19plus V2 with plumbing config

Strany 119

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 94 Replacing / Recharging Batteries Leave the batteries in place w

Strany 120

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 95 Recharging Optional Nickel Metal Hydride Batteries See the NiMH

Strany 121

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 96 4. Reinstall the battery pack in the housing: A. Align the bat

Strany 122

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 97 Recharging Optional Nickel-Cadmium Batteries The Ni-Cad battery

Strany 123

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 98 Conductivity Cell Maintenance The SBE 19plus V2’s conductivity

Strany 124

Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2 99 Pressure Sensor Maintenance Pressure sensor maintenance varies,