Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2
29
5. Display 19plus V2 status information by typing DS and pressing the Enter
key. The display looks like this:
SBE 19plus V 2.3 SERIAL NO. 4000 15 Mar 2011 14:02:13
vbatt = 10.1, vlith = 8.9, ioper = 61.9 ma, ipump = 20.8 ma,
iext01 = 76.2 ma
status = not logging
number of scans to average = 1
samples = 10, free = 4386532, casts = 1
mode = profile, minimum cond freq = 3000, pump delay = 60 sec
autorun = no, ignore magnetic switch = no
battery type = alkaline, battery cutoff = 7.5 volts
pressure sensor = strain gauge, range = 1000.0
SBE 38 = no, WETLABS = no, OPTODE = no, Gas Tension Device = no
Ext Volt 0 = yes, Ext Volt 1 = yes
Ext Volt 2 = no, Ext Volt 3 = no
Ext Volt 4 = no, Ext Volt 5 = no
echo characters = yes
output format = converted decimal
output salinity = no, output sound velocity = no
6. Command the 19plus V2 to take a sample by typing TS and pressing the
Enter key. The display looks like this (if in Profiling mode, with
converted decimal output format, no output salinity or sound velocity, and
auxiliary voltage sensors on channels 0 and 1):
23.7658, 0.00019, 0.062, 0.5632, 2.3748
where 23.7658 = temperature in degrees Celsius
0.00019 = conductivity in S/m
0.062 = pressure in db
0.5632 = voltage for auxiliary sensor channel 0
2.3748 = voltage for auxiliary sensor channel 1
These numbers should be reasonable; e.g., room temperature, zero
conductivity, barometric pressure (gauge pressure).
7. Command the 19plus V2 to go to sleep (quiescent state) by typing QS and
pressing the Enter key.
The 19plus V2 is ready for programming and deployment.
• The 19plus V2 automatically enters
quiescent (sleep) state after
2 minutes without receiving a
command. This timeout algorithm is
designed to conserve battery energy
if the user does not send QS to
put the 19plus V2 to sleep.
If the system does not appear to
respond, click Connect in the
Communications menu to
reestablish communications.
• Sending the status command
causes the pump to turn on for a
moment, so that the 19plus V2 can
measure and output the pump
current. Because the pump is
designed to be water lubricated, you
may hear a noise when the impeller
spins in air. Running the pump dry
for short periods (for example, when
sending the status command) will
not harm the pump.